DMSE Breakerspace - Trained-user reference
Nicolet iS5 FTIR Spectrometer
Thermo Scientific ยท Routine iD7 ATR SOP. The complete normal-use sequence in one place; use the OMNIC visual map on the reverse to locate controls.
Standard Operating Procedure (SOP)
1 Instrument startup
- Power on the instrument, if needed.
- Log on to the instrument workstation using your MIT Kerberos.
- Start OMNIC software.
- Verify System Status shows a green check. If it does not, stop and ask staff.
- Remove the protective cover from the ATR crystal plate.
- Clean and dry the ATR crystal.
- Collect a background before loading samples.
2 Operation
- Wear nitrile gloves when handling samples, ATR accessories, the pressure tower, pressure tips, Kimwipes, or cleaning solvent.
- Remove gloves before using the keyboard, mouse, or instrument workstation.
- Clean the ATR crystal before each sample or whenever contamination is possible.
- Choose the sampling method and pressure tip appropriate for the sample.
- Load the sample so it makes good contact with the ATR crystal or the selected accessory.
- Collect the sample spectrum.
- Save each spectrum you need. Spectra must be selected and saved individually.
- Wear gloves again before unloading the sample or cleaning the crystal.
- Repeat background, sample collection, cleaning, and saving as needed.
- If negative or unexpected features suggest a contaminated background, wear gloves to unload the sample, clean and dry the crystal and plate, then remove gloves and collect a complete new background with the crystal bare and unclamped. Reload and recollect the sample; stop and ask staff if the features remain.
3 Instrument shutdown
- Save all data you need and copy it to your own storage; the workstation is not a backup.
- Close OMNIC.
- Log off the workstation.
- Clean and dry the ATR crystal.
- Put the cover on the crystal plate and clamp it in place using the pressure tower.
- Leave the instrument powered on. The manufacturer recommends leaving the instrument powered when not in use.
- Leave the work area clean and remove all samples, wipes, and waste.
DMSE Breakerspace - Nicolet iS5 FTIR
OMNIC visual map
Use this side to find the controls. Follow the SOP on the reverse for the workflow.
Main OMNIC window
The routine controls stay in the top toolbar; System Status must show a green check.
Preview checks sample contact
Level 1 uses Preview Data Collection enabled. The live background preview is not a pass/fail test; confirm its clean, dry, bare, unclamped setup instead.