DMSE Breakerspace · Trained-user reference
Aeris Research X-Ray Diffractometer
Prepare a flat, level sample, run a stored program, and export both the raw scan and your processed result.
XRD quick reference
What XRD shows
Atomic planes in a crystal scatter X-rays, and at certain angles the scattered waves reinforce each other. The instrument records where those reflections appear across a range of angles (2θ). Peak positions come from atomic-plane spacings and are what identifies a phase — so anything that shifts them, especially sample height, costs you the result.
Select a measurement program
Choose a stored program from the drop-down list; you do not set scan parameters by hand. If none looks right for your sample, ask staff. Budget your reservation for every queued sample plus loading, export, and cleanup — a high-resolution program takes considerably longer than a quick phase-ID scan.
Choose the right holder
Powder, back or front loading → PW1811/00 or PW1811/27 · Odd shapes, fixed with plasticine or wax → PW1812/00 · Metal plates, membranes, pressed pellets → PW1813/26. Panalytical's prep guide in the lab lists every holder type.
Compatible samples
- Non-hazardous materials only. Powders, plates, pellets, coupons, membranes, and odd solids all have a holder.
- Grind powders to a few micrometers (roughly 5–10 µm), which also reduces preferred orientation.
- Grind gently, avoid airborne dust, wear safety glasses, and contain spills. Ask staff about fine, floaty, or irritating powders.
- Iron- or manganese-bearing samples fluoresce under Cu Kα, raising background and possibly preventing a usable result.
Before you leave the instrument
- Check that the pattern has peaks and reasonable counts, not a flat trace.
- Broad humps with no sharp peaks mean an amorphous material — a real result, not a failure.
- Unexpected extra peaks? Run an empty holder for comparison while you are still here.
- Weak, smeared, or shifted peaks usually mean sample height or grain size, both fixable by re-preparing now.
Background subtraction, peak search, and database matching happen afterward in HighScore Plus. See Data Processing And Analysis on the instrument page.
DMSE Breakerspace · Aeris Research XRD
Routine trained-user workflow
1 Prepare and start up
- Prepare your sample at the sample prep table using the correct holder. Holders and mounting clay are in the black cabinet beside the table.
- Load the holder so the sample surface is flat, level, centered, and flush with the reference surface.
- Label the holder physically. Plan a descriptive measurement name such as
Lastname_sampleID_YYYYMMDD. - Leave at least one changer position empty. Booting with all six occupied causes a startup error.
- Confirm the cover is closed and the interlocks are engaged. The mains switch at the rear is normally left on.
- Press the power button, then turn the HT keyswitch clockwise to switch on the high-tension generator.
2 Load and queue
- Remove the plastic sample-changer cover and place the holder in any free position. Note the slot number.
- Replace the plastic cover.
- In the software, select the changer position matching the physical slot. A mismatch scans the wrong slot or an empty one.
- Choose a measurement program from the drop-down list.
- Enter the sample name and edit the file name as needed.
- Start the measurement, or click Add to Queue to run it after the current sample.
- More samples can be loaded and queued while a scan runs. Confirm each one is actually added to the queue.
3 Export your data
- Export to the XRD workstation to the right of the instrument, or to a USB drive. Follow the current sign-in and export-folder instructions from your training or posted at the instrument.
- Export both the raw scan and any processed plots so you can reprocess later.
- Scans save in Panalytical's
.xrdmlformat; a plain-text or.csvpattern can also be exported for use elsewhere. - Keep a copy in your own storage. The shared workstation is not a backup.
- Phase matching in HighScore Plus runs on the lab workstation, which holds the reference database.
4 Log, clean, and shut down
- Confirm every file you need has been exported.
- Fill in the X-ray safety log book. This is the only record that you used the instrument.
- Press the power button to switch off the instrument.
- Turn the HT keyswitch counter-clockwise to switch off the HT generator.
- Remove your holders, return them to storage, and leave the sample prep area clean.
- Switch off the mains supply only if the instrument will be off for a long period.