DMSE Breakerspace - Trained-user reference

Malvern Panalytical Aeris Research XRD

Routine measurement SOP. The canonical normal-use sequence in one place; use the Aeris queue visual map on the reverse to locate controls taught in Level 1.

Provisional composition draft — incomplete

Standard Operating Procedure (SOP)

1 Instrument startup

  • Make sure at least one position in the sample changer is free (no sample holder in place). Booting with all six positions occupied causes a startup error.
  • Turn on the mains power switch at the rear of the instrument, if needed. This switch is normally left on.
  • Make sure the cover is closed and the interlocks are engaged.
  • Press the power button to switch on the instrument.
  • Turn the HT keyswitch clockwise to switch on the high-tension (HT) generator.

2 Operation

  • Prepare your sample at the sample prep table using the correct holder (see compatible materials and sample prep).
  • Remove the plastic sample-changer cover and place the prepared holder in any free loading position.
  • Replace the plastic sample-changer cover.
  • In the software, select the changer position that matches the physical slot where you placed the holder (see the queue interface).
  • Select a measurement program from the drop-down list.
  • Enter a descriptive sample name and edit the file name as needed.
  • Start the measurement, or click Add to Queue to run it after the current sample.
  • Export results to the networked workstation or a USB drive.
  • Fill out your information in the X-ray safety log book.

3 Instrument shutdown

  • Confirm all data you need has been exported.
  • Press the power button to switch off the instrument.
  • Turn the HT keyswitch counter-clockwise to switch off the HT generator.
  • Remove your sample holders, return them to storage, and leave the sample prep area clean.
Review gap

Decision required: Curate and document the routine stored-program list; until then retain the canonical direction to choose a stored program and ask staff if none looks right.

Content derived from the canonical Aeris Research XRD SOP - Owner: lab manager Provisional 2026-08-10 - Page 1 of 2

DMSE Breakerspace - Malvern Panalytical Aeris Research XRD

Aeris queue visual map

Use this side to find the queue controls taught in Level 1. Follow the SOP on the reverse for the canonical workflow.

Provisional composition draft — incomplete

Current Aeris queue screen

The approved map must show the routine queue controls in one current, sanitized native-resolution state.

Review gap

Decision required: Decide whether HighScore background/peak work and Search Match are independently taught Level 1 operations, a short capability preview, or a later analysis exercise. The provisional primary map covers only the Aeris queue.

Sample-changer cover readiness

Match the physical slot to the software position and restore the plastic cover before starting.

Stop: Do not start until the plastic sample-changer cover is replaced and the software position matches the physical slot.
Ready to queue: The holder is loaded, the plastic cover is restored, and the selected software position matches the physical slot.
Review gap

Decision required: Resolve companion-workstation sign-in and export/copy sequence, verify the queue field labels, and approve the changer-cover pair. This draft remains blocked for technical completion, publication, printing for use, and installation.